Eppendorf Xplorer® (plus) — Operating manual

22

If you have selected the Spc operating mode:

1. Use the rocker to select the required operating mode from the list.
2. Press the

Select

softkey.

5.5

Setting the parameter (edit mode)

Requirement

Pipette switched on (see p. 19)

Operating mode selected (see p. 21)

Change the parameters as follows:

1. Press the

Edit

softkey to open the edit mode.

2. Press the

Next

softkey to go to the next dispensing parameter.

3. Change the dispensing parameter marked in the display with the rocker.
4. Press the

End

softkey to save the changes, to exit the edit mode and to continue with

dispensing.

Only available for the Xplorer plus pipette

Hint!

If you want to cancel editing without saving the changes made, briefly turn the selection dial to
another mode.

Hint!

You can change the

Speed

,

Time

,

Cycles

and

Counter

parameters during the dispensing.

Press the

Edit

softkey during the execution.

Set volume

Aspiration and dispensing volume.
You can change the speed that the Xplorer pipette uses to change the
volume under the

Rocker setting

parameter in the options.

Set max. volume

Maximum aspiration and dispensing volume.
The set value appears after editing in the display next to

Max

.

Set volume/step

Volume per dispensing step. The maximum number of aspiration
steps (

Steps

) will be displayed when the

Set volume/step

is changed.

Aspiration speed

Aspiration speed
8 levels are available. The higher the numerical value, the higher the
speed.

Aspiration speed

will be automatically and approximately used for the

blow-out (

Blow

).

5

Operation