5.13.4 Reverse pipetting (Spc dial setting)

The

Rev. Pipetting

mode is well suited for plasma, serums and other liquids with a high protein

content. The

Pipetting

mode is also well suited for aqueous solutions (see p. 29). The

Rev.

Pipetting

mode is also well suited for solutions containing wetting agents, in order to minimize the

formation of foam during dispensing in the target tube. The liquid is aspirated using blow-out
(

Blow

). The blow-out is not part of the dispensing volume and may not be dispensed into the

target tube. If you use the same liquid again, the blow-out may remain in the tip. If you use a
different liquid, discard the blow-out and the pipette tip.

Requirement

Pipette switched on (see p. 19)

Rev. Pipetting

(Spc) operating mode selected (see p. 33)

Parameters set (see p. 22)

Pipette tip attached (see p. 24)

1. Press the rocker up to aspirate the blow-out and the liquid to be dispensed.

The aspiration of the dispensing volume and blow-out will be displayed under the dispensing
volume with the label

+ Blow 

.

2. Press the rocker down to dispense the liquid.

The piston returns to the basic position. Liquid remains in the tip after the dispensing volume
has been dispensed.

Press the rocker up to aspirate the same liquid.

Press the rocker down to end the pipetting and discard the blow-out (

Blow

).

Only available for the Xplorer plus pipette

Aspiration

Dispensing

Aspirate or empty

Hint!

If necessary, activate the

Counter

setting in the (see p. 42)options to count the executed

dispensings using the pedometer.

5

Operation