Eppendorf Xplorer® (plus) — Operating manual

48

5.17.12

Service

The options listed here are not needed during normal routine use. A selection list offers the
following service options:

Software version

: Display of the software that is valid for this Xplorer.

Parameter ver.

: Display of the applicable data record for this Xplorer. The data record is

different for the various volumes.

Initial reset

: After a security query, the pipette will be reset to factory settings and all entries

will be reset to the "default values". This type of reset may be useful when transferring the
Xplorer pipette to another workstation.

Check procedure

: For service personnel only. Not released for general use.

Software update

: For service personnel only.

USB

: For service personnel only.

5.17.13

Service interval

If the

Service interval

option is activated, the pipette will remind you about the required

maintenance or repeating the gravimetric test. You can set the reminder for a specific time or
after a specific number of cycles. The display shows the following options after

Time interval

or

Cycles interval

has been selected:

Last service

Next service

Set interval

Reset

For the

Set interval

parameter, set the required interval. If you would not like to use the

Service

interval

option, set the parameter value to

0

. The interval will be reset when the reset function is

triggered. If the pipette should start counting at a specific initial value, enter the value under

Set

interval

and then select

Reset

. You can view the last time a reset was triggered for the executed

maintenance under

Last service

You can view when the next maintenance needs to be executed

under

Next service

. If the interval has elapsed up to 90 %, the pipette shows the next required

maintenance in the display. Monitoring of the interval time and the interval cycles function
independently of each other. If you activate the interval time and interval cycles, you will receive a
maintenance reminder for both settings. A reset may be required after the maintenance and the
gravimetric test.

5.17.14

Date and Time

You can set the date and time of the pipette.

Select and change the parameter as described above (see p. 19).

Hint!

The

Initial reset

function can also be executed if the

Pswrd. protection

option is active. Please

contact Eppendorf AG Application Support if you would like to remove the password protection
without losing saved settings and programs. You will receive a

Master key

and additional

information on how to proceed.

Only available for the Xplorer plus pipette

Only available for the Xplorer plus pipette