12 service, 13 service interval, 14 date and time – Eppendorf XplorerPlus User Manual
Page 48: See p. 48), Eppendorf xplorer® (plus) — operating manual, Operation
Eppendorf Xplorer® (plus) — Operating manual
48
5.17.12
Service
The options listed here are not needed during normal routine use. A selection list offers the
following service options:
•
Software version
: Display of the software that is valid for this Xplorer.
•
Parameter ver.
: Display of the applicable data record for this Xplorer. The data record is
different for the various volumes.
•
Initial reset
: After a security query, the pipette will be reset to factory settings and all entries
will be reset to the "default values". This type of reset may be useful when transferring the
Xplorer pipette to another workstation.
•
Check procedure
: For service personnel only. Not released for general use.
•
Software update
: For service personnel only.
•
USB
: For service personnel only.
5.17.13
Service interval
If the
Service interval
option is activated, the pipette will remind you about the required
maintenance or repeating the gravimetric test. You can set the reminder for a specific time or
after a specific number of cycles. The display shows the following options after
Time interval
or
Cycles interval
has been selected:
•
Last service
•
Next service
•
Set interval
•
Reset
For the
Set interval
parameter, set the required interval. If you would not like to use the
Service
interval
option, set the parameter value to
0
. The interval will be reset when the reset function is
triggered. If the pipette should start counting at a specific initial value, enter the value under
Set
interval
and then select
Reset
. You can view the last time a reset was triggered for the executed
maintenance under
Last service
You can view when the next maintenance needs to be executed
under
Next service
. If the interval has elapsed up to 90 %, the pipette shows the next required
maintenance in the display. Monitoring of the interval time and the interval cycles function
independently of each other. If you activate the interval time and interval cycles, you will receive a
maintenance reminder for both settings. A reset may be required after the maintenance and the
gravimetric test.
5.17.14
Date and Time
You can set the date and time of the pipette.
Select and change the parameter as described above (see p. 19).
Hint!
The
Initial reset
function can also be executed if the
Pswrd. protection
option is active. Please
contact Eppendorf AG Application Support if you would like to remove the password protection
without losing saved settings and programs. You will receive a
Master key
and additional
information on how to proceed.
Only available for the Xplorer plus pipette
Only available for the Xplorer plus pipette
5
Operation