Technical data

9

Technical data

9.1

Eppendorf Xplorer single-channel

Model
Increment

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits Eppendorf AG

Error

Systematic error

Random error

± %

± μL

± %

± μL

0.5 - 10 μL
Increment:
0.01 μL

medium gray
0.1 - 20 μL
40 mm

1 μL

±2.5 

± 0.025

±1.8 

± 0.018

5 μL

±1.5 

±0.075 

±0.8 

±0.04 

10 μL

±1.0 

±0.1 

±0.4 

±0.04 

5 - 100 μL
Increment:
0.1 μL

yellow
2 - 200 μL
53 mm

10 μL

±2.0 

±0.2 

±1.0 

±0.1 

50 μL

±1.0 

±0.5 

±0.3 

±0.15 

100 μL

±0.8 

±0.8 

±0.2 

±0.2 

15 - 300 μL
Increment:
0.2 μL

orange
15 - 300 μL
55 mm

30 μL

±2.5 

±0.75 

± 0.7

± 0.21

150 μL

±1.0 

±1.5 

±0.3 

± 0.45

300 μL

±0.6 

±1.8 

±0.2 

±0.6 

50 - 1 000 μL
Increment:
1 μL

blue
50 - 1 000 μL
71 mm

100 μL

±3.0 

±3.0 

±0.6 

±0.6 

500 μL

±1.0 

±5.0 

±0.2 

± 1

1 000 μL

±0.6 

±6.0 

±0.2 

± 2

0.25 – 5 mL
Increment:
0.005 mL

purple
0.1 –5 mL
120 mm

0.5 mL

±3.0 

± 15.0

±0.6 

± 3

2.5 mL

±1.2 

±30.0 

±0.25 

± 6.25

5 mL

±0.6 

±30.0 

±0.15 

± 7.5

0.5 – 10 mL
Increment:
0.01 mL

turquoise
1 –10 mL
165 mm

1 mL

±3.0 

±30.0 

± 0.60

±6.0 

5 mL

±0.8 

± 40.0

± 0.20

± 10.0

10 mL

±0.6 

± 60.0

±0.15 

± 15.0

Hint!

You can select a volume of up to 0.20 mL for the 5 mL Xplorer pipette. For the 0.2 mL – 0.5 mL
volume range, handling has a strong influence on the achievable random errors.