10 language, 11 personalization, 10 language 5.17.11 personalization – Eppendorf XplorerPlus User Manual
Page 47: Ord (see pswrd. protection on p. 47), Eppendorf xplorer® (plus) — operating manual, Operation
![background image](/manuals/551179/47/background.png)
47
Eppendorf Xplorer® (plus) — Operating manual
•
3-point adjust.
After you enter the density, three different volumes and the corresponding weighing results,
the Xplorer pipette determines two correction factors. The factors are correct from measuring
point to measuring point in the selected volume regions and for the selected work technique.
The respective factor is, however, also used below and above the first or third measuring
point. You must carry out a gravimetric test to check the validity of the factors.
Pswrd. protection
For the Xplorer plus, the adjustment can be protected with a four-digit numeric password. The
passwords for the
Edit
mode and
Adjustment
option may vary. The password can be changed or
deactivated after it has been entered. Please contact Eppendorf AG Application Support or
conduct a reset if you lose the password.
1. Press the rocker upward to activate the password protection.
ON
appears in the display.
2. Press the
Select
softkey.
3. Use the rocker to select the digit.
4. Press the
Next
softkey to switch to the next field.
5. Enter data in all of the fields.
6. Press the
Save
softkey to save the password.
7. Reenter the password after completing the saving procedure. Press the
Enter
softkey.
Password protection is now active.
5.17.10
Language
Mark a language in the selection list with the rocker.
The selected language is only activated after the completion of the language selection.
5.17.11
Personalization
You can use the rocker to select letters and numbers to personalize the pipette. The selected
personalization will be displayed after the pipette has been activated from sleep mode or after the
reset. When the Xplorer pipette is delivered, it is personalized as
My Xplorer
.
Hint!
You must gravimetrically check each Xplorer pipette whose factory setting has been changed by
selection or by performing the above-mentioned adjustments. This is the only way to ensure that
the selected adjustment meets requirements regarding errors.
Only available for the Xplorer plus pipette
Hint!
The adjustment can no longer be changed if the password has been lost.
Hint!
When selecting a language, the softkeys and the language selection list are always in English.
This makes it easier to return to the desired language if you select an unknown language by
mistake.
5