2 eppendorf xplorer multi-channel, Eppendorf xplorer multi-channel, Eppendorf xplorer® (plus) — operating manual – Eppendorf XplorerPlus User Manual
Page 64: Technical data
Eppendorf Xplorer® (plus) — Operating manual
64
9.2
Eppendorf Xplorer multi-channel
The technical data applies to both the 8-channel and 12-channel pipettes.
Model
Increment
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Error limits Eppendorf AG
Error
Systematic error
Random error
± %
± μL
± %
± μL
0.5 - 10 μL
Increment:
0.01 μL
medium gray
0.1 - 20 μL
40 mm
1 μL
±5.0
± 0.05
±3.0
± 0.03
5 μL
±3.0
±0.15
±1.5
±0.075
10 μL
±2.0
±0.2
±0.8
± 0.08
5 - 100 μL
Increment:
0.1 μL
yellow
2 - 200 μL
53 mm
10 μL
±2.0
±0.2
±2.0
±0.2
50 μL
±1.0
±0.5
±0.8
±0.4
100 μL
±0.8
±0.8
±0.25
±0.25
15 - 300 μL
Increment:
0.2 μL
orange
15 - 300 μL
55 mm
30 μL
±2.5
±0.75
±1.0
±0.3
150 μL
±1.0
±1.5
±0.5
±0.75
300 μL
±0.6
±1.8
±0.25
±0.75
50 - 1 200 μL
Increment:
1 μL
green
50 - 1 250 μL
76 mm
120 μL
±6.0
± 7.2
± 0.9
± 1.08
600 μL
± 2.7
± 16.2
±0.4
± 2.4
1 200 μL
±1.2
± 14.2
±0.3
± 3.6
9
Technical data