Eppendorf Xplorer® (plus) — Operating manual

Dispensing speed

Dispensing speed
8 levels are available. The higher the numerical value, the higher the
speed.

Dispensing speed

will be automatically and approximately used for

the blow-out (Blow).

Set time

Interval between the dispensing step (0.1 s to 10 s).

Number of steps

Number of dispensing steps
The available number of dispensing steps depends on the setting for

Set volume/step

and is automatically set to the maximum number

when

Set volume/step

is edited.

Under steps

in the "Sequential dispensing" mode, the selected

number of dispensing steps will be displayed to the right, and the
dispensing step for the displayed volume will be shown at the left.

Set counter

The counter counts the number of dispensing operations performed.
The numerical value of the counter is increased by

1

when the piston

reaches the basic position after dispensing.
If the piston is in the basic position, you can alternatively set the
counter to

0

by turning the selection dial forward or backward.

You can switch the counter on and off in the Options.

Mixing volume

The mixing volume used after pipetting. If you change the pipetting
volume under

Set volume

, the mixing volume changes accordingly.

Mixing cycles
1

to

99

mixing cycles can be set for the mixing volume. The mixing

cycles are counted down to

0

. If you keep the rocker pressed down

during mixing, after

0

has been reached, mixing is continued until the

rocker is released. If you press the rocker up during mixing, the mixing
process is stopped.
The mixing speed matches the selected speed for aspiration and
dispensing.

Set sample vol.

Volume of the sample or reagent
The maximum potential diluent volume will be automatically set if the
volume of the sample is changed.

5

Operation