Eppendorf Xplorer® (plus) — Operating manual

34

5.13.2 Diluting (Spc dial setting)

Two different liquids are aspirated in the

Diluting

mode. These liquids are separated by an air

bubble. The

Diluting

mode is suitable for diluting samples and reagents using an appropriate

dilution solution (diluent). First, the diluent is aspirated, then the air bubble, and finally the sample
or reagent.

Requirement

Pipette switched on (see p. 19)

Diluting

(Spc) operating mode selected (see p. 33)

Parameters set (see p. 22)

Pipette tip attached (see p. 24)

1. Press the rocker up to aspirate the diluent.
2. Press the rocker up to aspirate the air bubble.
3. Press the rocker up to aspirate the sample or reagent.
4. Press the rocker down to dispense all of the liquid.
5. Press the rocker down to trigger the blow-out

Blow

.

6. If the piston is in the basic position: press the

Back

softkey to exit the mode.

Examples:

Required auxiliary aids:

Xplorer plus with a nominal volume of 1 000 μL

Dispensing tube (1.5 mL)

Only available for the Xplorer plus pipette

Hint!

Because both liquids are aspirated in a single pipette tip, the sample or reagent may be slightly
contaminated by the dilution solution.

Diluent

Aspirating the air

bubble

Aspiration of the

sample or reagent

Dispensing

Hint!

For execution, the parameter sequence is carried out in the reverse order of the editing process.
The next volume to be aspirated will be shown in the display with a border around it.

Hint!

Set a high value for the

Dispensing speed

parameter to achieve a good mixture for the

dispensing.

Hint!

To create a very high mixture, you can enter the "Diluting" mode with the Pipetting and Mixing
mode (P/M) as the program (see p. 39).