Selecting the operating mode

You can set the desired operating mode using the selection dial.
The following operating modes are available:

Selection dial Mode

Xplorer

Xplorer plus

Ads

Auto dispensing
Dispensing partial volumes in a defined
interval

•

•

Dis

Dispensing
Dispensing uniform partial volumes

•

•

Pip

Pipetting
Aspiration and dispensing of liquid

•

•

P/M

Pipetting and Mixing
Pipetting with freely selected mixing volume

•

•

Man

Manual pipetting
Aspiration and dispensing when the rocker
is held pressed down

•

•

Opt

Options
Making settings and changing the
adjustment

•

•

Spc

Multiple aspiration
Aspiration of uniform partial volumes and
combined dispensing (reverse dispensing)

•

Diluting
For diluting a sample or reagent

•

Sequential dispensing
Functions like dispensing mode with a fixed
sequence of various volumes

•

Reverse pipetting
Functions like the pipetting mode, but in
reverse order

•

Sequential pipetting
For creating pipetting series with various
volumes

•

Prg

Program
Execute saved dispensings

•

Edit

Editing
Procedure programs and pipettings with
fixed volumes

•

Fix

Pipetting with fixed volumes
Procedure pipettings with fixed volumes.

•

To select an operating mode, turn the selection dial to the selected mode.