5.13.3 Sequential dispensing (Spc dial setting)

Up to 10 different volumes can be dispensed in the

Seq. Dispensing

mode. The sum of the

dispensing volumes may not exceed the pipette tip filling volume. This mode is well suited for
dilution series. The

Seq. Pipetting

mode is also well suited for dilution series. To conduct a

dilution series, dispense two different liquids in the opposite direction. This allows you to achieve
different dilutions with the same volume per tube.

Requirement

Pipette switched on (see p. 19)

Seq. Dispensing

(Spc) operating mode selected (see p. 33)

Parameters set (see p. 22)

Pipette tip attached (see p. 24)

1. Press the rocker up to aspirate liquid.
2. Press the rocker down to execute the reverse stroke.
3. Press the rocker down for each dispensing step.
4. If the piston is in the basic position: press the

Back

softkey to exit the mode.

Example

Required auxiliary aids:

Xplorer plus with a nominal volume of 300 μL

4 dispensing tubes (200 μL)

Only available for the Xplorer plus pipette

Hint!

If the sum of the dispensing volumes exceeds the volume of the pipette tip: create a program and
execute the mode several times in a row.

Tube 1

Tube 2

Tube 3

Tube 4

Volume/step

30 μL

60 μL

90 μL

120 μL

Volume/step

120 μL

90 μL

60 μL

30 μL

Dilution 1+X

1+4

1+1.5

1+0.67

1+0.25

Dilution 1:Y

1:5

1:2.5

1:1.67

1:1.25

Tube sum

150 μL

150 μL

150 μL

150 μL

1

2

3

4

6

5

1

1

2

3

4

6

5

2

5

Operation