5 sequential pipetting (spc dial setting), Sequential pipetting (spc dial setting), Eppendorf xplorer® (plus) — operating manual – Eppendorf XplorerPlus User Manual
Page 38: Operation
Eppendorf Xplorer® (plus) — Operating manual
38
5.13.5 Sequential pipetting (Spc dial setting)
In the
Seq. Pipetting
mode, a maximum of 10 different pipetting volumes are dispensed in a fixed
sequence. This mode is well suited for dilution series. The
Seq. Dispensing
mode is also well
suited for dilution series. The random error (precision) and systematic error (accuracy) are better
for pipetting than with dispensing. The
Seq. Pipetting
mode offers more flexibility with the volume
selection than the
Seq. Dispensing
mode.
Requirement
•
Pipette switched on (see p. 19)
•
Seq. Pipetting
(Spc) operating mode selected (see p. 33)
•
Parameters set (see p. 22)
•
Pipette tip attached (see p. 24)
1. Press the rocker up to aspirate liquid.
2. Press the rocker down to dispense the liquid.
3. Press the rocker downward after the dispensing to execute the blow-out (
Blow
).
4. If the piston is in the basic position: press the
Back
softkey to exit the mode.
Only available for the Xplorer plus pipette
Aspiration
Dispensing
Aspiration
Dispensing
Hint!
After each dispensing, you can repeat the blow-out (
Blow
) as many times as you would like by
pressing the rocker down.
Observe the information on the blow-out (
Blow
5
Operation