beautypg.com

5 sequential pipetting (spc dial setting), Sequential pipetting (spc dial setting), Eppendorf xplorer® (plus) — operating manual – Eppendorf XplorerPlus User Manual

Page 38: Operation

background image

Eppendorf Xplorer® (plus) — Operating manual

38

5.13.5 Sequential pipetting (Spc dial setting)

In the

Seq. Pipetting

mode, a maximum of 10 different pipetting volumes are dispensed in a fixed

sequence. This mode is well suited for dilution series. The

Seq. Dispensing

mode is also well

suited for dilution series. The random error (precision) and systematic error (accuracy) are better
for pipetting than with dispensing. The

Seq. Pipetting

mode offers more flexibility with the volume

selection than the

Seq. Dispensing

mode.

Requirement

Pipette switched on (see p. 19)

Seq. Pipetting

(Spc) operating mode selected (see p. 33)

Parameters set (see p. 22)

Pipette tip attached (see p. 24)

1. Press the rocker up to aspirate liquid.
2. Press the rocker down to dispense the liquid.
3. Press the rocker downward after the dispensing to execute the blow-out (

Blow

).

4. If the piston is in the basic position: press the

Back

softkey to exit the mode.

Only available for the Xplorer plus pipette

Aspiration

Dispensing

Aspiration

Dispensing

Hint!

After each dispensing, you can repeat the blow-out (

Blow

) as many times as you would like by

pressing the rocker down.

Observe the information on the blow-out (

Blow

) (see p. 29).

5

Operation

This manual is related to the following products: