5: Troubleshooting

47

SET-UP ERRORS

Pipe OD out of range

Interpretation: You have entered an out-of-range value for the pipe outside diameter
dimension – i.e. larger or smaller than the unit or sensor can be used on.

Response: Enter a valid number.

Wall thickness out of
range

Interpretation: You have entered an out-of-range value for the pipe wall thickness
dimension – accepted range is 0.04 - 3.0 inches (1 to 75 mm).

Response: Enter a valid number.

Lining thickness out of
range

Interpretation: You have entered an out-of-range value for the lining thickness
dimension – acceptable range is 0 - 0.4 inches (0 to 25 mm).

Response: Enter a valid number.

Temperature range

Interpretation: You have entered an out-of-range value for the fluid temperature.
Accepted temperature range -4°F to +572°F ( -20°C to +300°C).

Response: Enter a valid number.

Invalid Date or Time

Interpretation: The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.

Response: Enter a valid Date and Time.

Sensors: INVALID

Interpretation: The selected temperature is higher than the maximum allowed for the
sensor type.

Response: Select alternative sensors or change the temperature.

Mode: Err Typ

Interpretation: The selected sensors are invalid and the mode cannot be verified.

Response: Select a valid sensor type and choose a mode that gives a non-zero
separation distance.