3: Operating Procedures
18
Note: The
SENSOR SELECTION menu can be entered by use of the Up/Down scroll keys. This allows the
type and mode of the sensors being used to be selected. This menu will be entered automatically if the
entered pipe OD and/or temperature are not valid for the currently selected sensors.
Note: Do not press ENTER until the transducers are fitted and connected to the instrument.
Attaching and connecting the transducers
14. Fit the designated sensors to the pipe using the appropriate transducer holders as described in Paragraph
2.2. Take great care to set the separation distance as accurately as possible.
15. Connect the red and blue coaxial cables between the sensors and the instrument, ensuring that the red
connector on the instrument is connected to the ‘upstream’ sensor.
Taking a flow reading
7.
If you need to alter the fluid temperature from
that shown, select either
°C or °F with the cursor
and press the
ENTER key.
8.
Enter the new temperature value and press the
ENTER key.
9.
The new temperature should now be indicated in
both °C and °F.
10. Select
Continue.. and press the ENTER key.
FLUID TEMPERATURE
DD-MM-YY HH:MM:SS
Enter Fluid Temperature
°C: 5.00
°F: 41.00
Continue..
11. The
SENSOR SEPARATION screen now displays
a summary of the entered parameters and
informs you of the type of sensor to be used, the
mode of operation, and the distance to set up
between the sensors.
In this example, it recommends type
A-ST (A
standard) sensors operating in the ‘
Reflex’
mode spaced
3.62 Inches apart.
Take a note of these details.
SENSOR SEPARATION
DD-MM-YY HH:MM:SS
Site : MySite
Pipe : 6.50 Inches
Wall : 0.05 Inches
Sensors : B-ST Reflex
Temperature : 10.00°C 50.00°F
Set sensor separation to 3.62 Inches
Press
to continue,
UV
to select sens.
12. Take a note of the displayed parameters, then
press the
ENTER key.
13. The
ATTACH SENSORS screen displays, giving
instructions to attach the sensors.
ATTACH SENSORS
DD-MM-YY HH:MM:SS
Attach Sensor Set
Red Connector UpStream
Press
key to go to Flow Reading
16. Once the transducers have been fitted and
connected, press the
ENTER key on the ATTACH
SENSORS screen.
17. This will take you to the
FLOW READING screen
via a signal-checking screen (shown here).
Please wait..
Checking signals
****************************************
* *
****************************************