3: Operating Procedures

14

3.2

Using the Quick Start Menu

If you want to perform a ‘one time’ flow reading at a particular pipe location, the

Quick Start menu provides the

quickest way to set up the PUF system and access the

FLOW READING screen.

If the point at which you intend to take the measurement is likely to require regular monitoring, it is best to set it up
as a ‘Site’ within the PUF, which then stores the site parameters (See Paragraph 3.4).

Before you can use the PUF system you need to obtain the following details (this information will be required
when setting up the

Quick Start menu):

The pipe outside diameter.
The pipe wall thickness and material.
The pipe lining thickness and material.
The type of fluid.
The fluid temperature.

Entering the site data

1.

Select

Quick Start from the MAIN MENU and press the ENTER key. You will then be presented with a

series of screens in which to enter the data mentioned above.

2.

Select the dimension units (millimeters or inches)
used to measure the pipe, then press the

ENTER

key.

DIMENSION UNIT

DD-MM-YY HH:MM:SS

Select the dimension units:

mm

Inches

3.

Enter the pipe outside diameter dimension, then
press the

ENTER key.

OUTSIDE DIAMETER

DD-MM-YY HH:MM:SS

Dimension: Inches

Pipe outside diameter? 6.50

4.

Enter the pipe wall thickness dimension, then
press the

ENTER key.

PIPE WALL THICKNESS

DD-MM-YY HH:MM:SS

Dimension: Inches

Pipe outside diameter? 6.50

Pipe wall thickness? 0.50

5.

If the pipe has a lining, enter the lining thickness.
If nothing is entered the instrument automatically
assumes there is no lining.

6.

Press the

ENTER key to continue.

PIPE LINING THICKNESS

DD-MM-YY HH:MM:SS

Dimension: Inches

Pipe outside diameter? 58.0

Pipe wall thickness? 4.0

Pipe lining thickness? 0.12