Flow monitoring, Ph 3.3), Operating procedures – Dwyer Series PUF User Manual
Page 21
3: Operating Procedures
17
Flow monitoring
The
FLOW READING screen is the one most used during normal monitoring operation. It shows the instantaneous
fluid flow together with totalized values (when enabled). In this mode you can select the flow rate measurement
units by pressing keys 7 (liters), 8 (Gallons, Barrels) or 9 (m³), or change the display to show velocity by pressing
key 4.
If the flow reading exceeds a value of +/-9999 in the current units, a *10 multiplier will be displayed above the
units and the value displayed will be a tenth of the actual value. Similarly a * 100 and *1000 may be displayed on
very large flow rates.
There are limitations on the use of these larger flow values with regards to logging of data and setting the current
and pulse outputs. If the flow reading is less than 140000 US gallons/min, the value can be logged and
downloaded via Portagraph III.
3.3
Using the System at a Regularly Monitored Location
Setting up the PUF system using the
Quick Start method described in Paragraph 3.2 is easy and the
recommended method to use in a ‘one time’ situation. But if you have a site location that you want to monitor on a
frequent basis it is better to set up a named ‘Site’ for that location so that you can recall it when needed and so
avoid the need to re-enter the site details every time you want to install the equipment there.
Note: See Paragraph 3.4 for details of how to set-up and manage site details.
Use this procedure to install the equipment at a named site.
Note: If you select a different Sensor set (e.g. A-ST) when entering new site data you could receive an “invalid”
error message if the previous Sensor set was operating at a temperature above 275ºF. If this occurs, ignore the
warning as it will disappear when you enter a temperature in the correct range for new sensors.
1.
Select
View Edit Site Data from the MAIN
MENU.
2.
Select
Choose from list of sites.
3.
Select one of the sites listed and press the
ENTER key.
4.
The
Site name will show the selected site and
the site parameters will be listed on the screen.
5.
Scroll down through the menu list and enter/
change the data that might have changed since
the last time the site was accessed.
6.
When you are satisfied that the parameters are
correct select
Save current site & read
flow.
VIEW EDIT SITE DATA
DD-MM-YY HH:MM:SS
Choose from list of sites
Site name : MySite
Dimension units : Inches
Pipe outside diameter: 6.50
Pipe wall thickness : 0.50
Pipe lining thickness: 0.16
Pipe wall material : Mild Steel
Lining material : Steel
Sensor set : B-ST
(see note)
Sensor mode : Reflex
Fluid type : Water
Save current site & read flow
Delete this Site
Download & save current site
Exit