3: Operating Procedures

20

3.4.2 Changing a site name

To change a site name, use the same method described above for generating a new site: but in this case access
a current site name to change rather than an EmptySite. If you change a site name while the site is logging, the
logging will stop.

3.5

Instrument Calibration

The PUF is fully calibrated before it leaves the factory; however the following adjustments are provided to allow
you to further ‘fine tune’ your instrument to suit local conditions and application where necessary. Apart from the
zero flow offset adjustment, these are normally carried out only where the instrument is to be used in a permanent
or semi-permanent location.

3.5.1 Adjusting the zero cut-off

This adjustment allows you to set a minimum flow rate (m/s) below which the instrument will indicate ‘0’. The
default setting is 0.1 m/s (0.33 ft/s) but you may adjust this value if required.

1.

With the instrument operating in

FLOW READING mode, press the Options key to access the FLOW

READING OPTIONS menu shown.

8.

Scroll down through the menu list and enter/
change the pipe parameters and other data
pertaining to the site.
Note that this menu allows you to choose a
Sensor Set, unlike the QUICK START menu
which recommended the sensor set to use. If
you enter an inappropriate sensor set in this
menu, you will be presented with an error
message later when you go to the

SENSOR

SEPARATION screen.

VIEW EDIT SITE DATA

DD-MM-YY HH:MM:SS

Choose from list of sites

Site name : MyNewSite

Dimensions : Inches

Pipe outside diameter: 0.00

Pipe wall thickness : 0.00

Pipe lining thickness: 0.00

9.

When all the data is correct you can either:

a)

Select

Save current site & read

flow to continue fitting the transducers
and opening the

FLOW READING screen.

b)

Select

Delete this site to delete the

site name and values and restore it to the
original

EmptySite name.

c)

Select

Download & save current

site to save the site details and download
them to a PC via RS232/USB provided the
RS232/USB link is connected and
correctly set-up.

d)

Select

Exit to return to the MAIN MENU.

Pipe wall material : Mild Steel

Lining material : ---------

Sensor set : A-ST

Sensor mode : Reflex

Fluid type : Water

Save current site & read flow

Delete this Site

Download & save current site

Exit

2.

Select

Zero Cutoff (m/s) and press the

ENTER key.

3.

Enter the value for the

Zero Cutoff (e.g. 0.06

m/s) then press the

ENTER key.

4.

Scroll down to select

Exit and press the ENTER

key to return to the

FLOW READING screen.

FLOW READING OPTION

DD-MM-YY HH:MM:SS

Data review

Zero Cutoff (m/s) : 0.00

Set zero flow (m/s) : 0.00

Damping (secs) : 10

Totalizer : Run

Reset +Total