3: Operating Procedures

39

Calibrating the pulse frequency range

The

Calculated Pulse Value field is automatically calculated as:

This value is often required to be programmed into the device that is receiving the pulses to enable it to calculate
the correct flow rate – for example, when connecting the pulse output to an energy meter.

3.9

Printing

Logged data can be printed using an RS232-compatible printer.

3.9.1 How to print logged data using the RS232 printer

This procedure shows you how to set up and use the RS232 printer to print logged data.

Setting up an RS232 connection

1.

Connect the PUF to the printer using the printer cable supplied. This cable is fitted with a 9-way D-Type plug
terminated as follows:

4.

Ensure that the printer has sufficient paper, is switched

ON and ON LINE.

1.

With the instrument operating in the

FLOW

READING mode, press the Pulse function key
to access the

PULSE OUTPUT screen.

2.

Select

Max Pulse Freq (Hz) and enter the

required value.

3.

Select

Flow at Max Freq and enter the

required value (gal/min).

4.

Select

Exit and press the ENTER key to return

to the

FLOW READING screen

PULSE OUTPUT

DD-MM-YY HH:MM:SS

A1 Pulse Frequency is ON

Qxx.xx%

Flow units : gal

Output : Frequency

Vol per pulse : 10.00

Pulse width (ms) : 10

Max Pulse Freq (Hz) : 10.00

Flow at Max Freq : 200.00

Calculated Pulse Value: 20.0

Exit

Pin 1 – N/C (No connection)
Pin 3 – RxD data from printer to PUF
Pin 5 – Signal ground
Pin 7 – N/C
Pin 9 – N/C

Pin 2 – TxD data from PUF to printer
Pin 4 – N/C
Pin 6 – N/C
Pin 8 – N/C

2.

Select

Setup RS232 / USB from the MAIN

MENU. This will access the SETUP RS232/USB
screen shown here.

3.

Set the RS232 parameters to comply with those
for the printer.

SETUP RS232 / USB

DD-MM-YY HH:MM:SS

Qxx.xx%

Handshaking : Xon/Xoff

Baud Rate : 1200

Data Bits : 8

Stop Bits : 1

Parity : None

New line : CR+LF

Printer test :

Exit

Calculated Pulse Value

Flow at Max Freq

Max Pulse Freq (Hz)

--------------------------------------------------

=