beautypg.com

Automatic signal loss recovery, 2 general troubleshooting procedure, Troubleshooting – Dwyer Series PUF User Manual

Page 48: Figure 5.1 troubleshooting chart

background image

5: Troubleshooting

44

Automatic signal loss recovery

If the signal is lost or the Quality falls below 40%, the set up procedure, normally invoked by

Read Flow in the

main menu, is automatically run until a good quality signal is found.

5.2

General Troubleshooting Procedure

Figure 5.1 Troubleshooting chart

Is the display blank?

Recharge the battery.

If battery won't recharge then

replace the charger (if faulty) or

return the instrument for repair.

No

Turn instrument OFF/ON.

If the display is still scrambled/

hung up press the microprocessor

reset button. Return instrument

for repair if fault still present.

Yes

Is the display

scrambled or

hung up?

Yes

Measurement differs from expected value.

Unstable measurements.

No (or poor) signal.

Problem probably due to non-constant fluid flow.

Check the following:

- the pipe data has been entered correctly.

- the fluid type has been entered correctly.

- the correct transducer type has been selected.

- the pipe diameter is within specifications of the
selected transducers.

- the pipe is completely full.

- the pipe surface is not corroded, or protective
surface loose.

- no particles in the fluid.

Ensure the temperature is set correctly.

Ensure that the transducer cables are connected

correctly.

Ensure transducers are positioned in accordance

with the recommended distance from bends etc.

Ensure flow velocity >0.1m/s.

Ensure that sufficient acoustic couplant has been

applied to the transducers.

If triple reflex mode is selected, try selecting

double or single reflex instead.

Check the instrument using the test block.

No

START

Are any

status messages

shown?

No

Note any active messages.

Refer to the message table in this

chapter to interpret the message

and view suggested response.

Yes