3: Operating Procedures
23
With the system running in
FLOW READING mode:
3.5.5 Adjusting the damping factor
By averaging-out the flow rate over several seconds, the
Damping factor can be used to smooth out rapid
changes in flow rate to prevent wild fluctuations in the displayed flow value. It has a range of 1, 10, 15, 20, 30, 50,
seconds, with a default setting of 10.
With the system running in
FLOW READNG mode:
1.
Press the
Options key to access the FLOW
READING OPTION screen shown.
2.
Scroll down and select
Roughness factor.
3.
Change the roughness factor according to the
pipe material and condition as described above.
4.
Press the
ENTER key to apply the change.
FLOW READING OPTION
DD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s) : 0.00
Set zero flow (m/s) : 0.00
Damping (secs) : 10
Totalizer : Run
Reset +Total
Reset –Total
Calibration factor : 1.000
Roughness factor : 0.01
Diagnostics
Select Totals : Both
Exit
1.
Press the
Options key to access the FLOW
READING OPTION screen shown.
2.
Scroll down and select
Damping (secs).
3.
This will open the
DAMPING OPTIONS screen.
FLOW READING OPTION
DD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s) : 0.00
Set zero flow (m/s) : 0.00
Damping (secs) : 10
Totalizer : Run
Reset +Total
Reset –Total
Calibration factor : 1.000
Roughness factor : 0.01
Diagnostics
Select Totals : Both
Exit
4.
Select the value of the
Damping factor as
required to remove any unwanted display
fluctuations. Increasing the value applies a
greater smoothing affect.
5.
Press the
ENTER key to apply the selection and
return to the
FLOW READING OPTIONS screen.
DAMPING OPTIONS
DD-MM-YY HH:MM:SS
1 second
10 seconds
15 seconds
20 seconds
30 seconds
50 seconds
Key Point: If the damping factor is set too high, the value displayed may appear
stable but it may exhibit large step changes when the value is updated.