beautypg.com

B.106 system:ramp command, B.107 system:ramp? query, B.108 system:ramp:positive command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual

Page 89: B.109 system:ramp:positive? query, B.110 system:ramp:negative command, B.111 system:ramp:negative? query, B.106, System:ramp command, B.107, System:ramp? query

background image

SERIES EL 071414

B-21

B.106 SYSTem:RAMP COMMAND

SYST:RAMP

Syntax:

Short Form: SYST:RAMP

Long Form: SYSTem:RAMP

Where

Description: Establishes the time for unit to change states in both positive and negative directions. See

SYST:RAMP:POS (PAR. B.108) and SYST:RAMP:NEG (PAR. B.110) to set different ramp times for
positive and negative directions.

Ramps work in all modes except power. The ramp consists of up to 4000 evenly spaced steps in times
from 4.5 microseconds to 2.50 milliseconds (max ramp time is 10 seconds). The ramp will either step
from one level to the other in 4.5 microsecond steps or will execute approximately 4000 steps at a time
required to meet the ramp duration. For example, a ramp of 1 millisecond consists of 222 steps (4.5
microseconds per step). A ramp of 100 milliseconds consists of 4000 steps (25 microseconds per
step). A ramp of 2 seconds consists of 4000 steps (0.5 milliseconds per step).

When using both ramps and transients at the same time, the following must be observed (see Figure
B-1 for timing, see PAR. B.100 to initiate a transient pulse.):
- Transient must be greater than set point if positive ramp and negative ramp are different.
- Transient pulse period

resistance and PAR. B.62 for voltage.

B.107 SYSTem:RAMP? QUERY

SYST:RAMP?

Syntax:

Short Form: SYST:RAMP?

Long Form: SYSTem:RAMP?

Returns:

Description: Indicates the larger of the positive and negative ramp times. For example, If SYST:RAMP:POS is

set to 2 and SYST:RAMP:NEG is set to 1.5, SYST:RAMP? returns 2.000.

B.108 SYSTem:RAMP:POSitive COMMAND

SYST:RAMP:POS

Syntax:

Short Form: SYST:RAMP:POS

Long Form: SYSTem:RAMP:POSitive

Where

Description: Establishes the time for unit to change states in positive direction. See SYST:RAMP command,

PAR. B.106 for ramp details. and restrictions.

B.109 SYSTem:RAMP:POSitive? QUERY

SYST:RAMP:POS?

Syntax:

Short Form: SYST:RAMP:POS?

Long Form: SYSTem:RAMP:POSitive?

Returns:

Description: Indicates the positive ramp time.

B.110 SYSTem:RAMP:NEGative COMMAND

SYST:RAMP:NEG

Syntax:

Short Form: SYST:RAMP:NEG

Description: Establishes the time for unit to change states in negative direction. See SYST:RAMP command,

PAR. B.106 for details and restrictions.

B.111 SYSTem:RAMP:NEGative? QUERY

SYST:RAMP:NEG?

Syntax:

Short Form: SYST:RAMP:NEG?

Long Form: SYSTem:RAMP:NEGative?

Returns: