beautypg.com

B.81 status:questionable:enable? query, B.82 system:communicate[:gpib]:address command, B.83 system:communicate[:gpib]:address? query – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual

Page 83: B.84 system:communicate:echo command, B.85 system:communicate:echo? query, B.86 system:communicate:lan:ipaddress command, B.87 system:communicate:lan:ipaddress? query, B.81, Statu, B.82

background image

SERIES EL 071414

B-15

B.81

STATus:QUEStionable:ENABle? QUERY

STAT:QUES:ENAB?

Syntax:

Short Form: STAT:QUES:ENAB?

Long Form: STATus:QUESionable:ENABle?

Return Value: =16 bits = actual Questionable enable register value

Description: Reads Questionable Condition Enable Register. Electronic load returns value of Questionable

Condition Enable Register, indicating which conditions are being monitored. Bit set to 1 = function
enabled (active, true); bit reset to 0 = function disabled (inactive, false)

.

B.82

SYSTem:COMMunicate[:GPIB]:ADDRess COMMAND

SYST:COMM[:GPIB]:ADDR

Syntax:

Short Form: SYST:COMM[:GPIB]:ADDR
Long Form: SYSTem:COMMunicate[:GPIB]:ADDRess
Where: = numbers from 1 to 30 comprising the GPIB Address. Factory default GPIB address
is 10.

Description: Sets the GPIB Address to be used during communication over the GPIB interface. After setting

the address, the EL Load must be power cycled (turned off, then on) in order for the new GPIB
address to take effect. Command is not valid from the GPIB interface.

B.83

SYSTem:COMMunicate[:GPIB]:ADDRess? QUERY

SYST:COMM[:GPIB]:ADDR?

Syntax:

Short Form: SYST:COMM[:GPIB]:ADDR?
Long Form: SYSTem:COMMunicate[:GPIB]:ADDRess?
Returns: = number from 1 to 30 comprising the GPIB Address

Description: Returns the GPIB Address used during communication over the GPIB interface. Command is

not valid from the GPIB interface.

B.84

SYSTem:COMMunicate:ECHO COMMAND

SYST:COMM:ECHO

Syntax:

Short Form: SYST:COMM:ECHO

Long Form: SYSTem:COMMunicate:ECHO

Where: 1 or ON turns on echo, 0 or off turns off echo

Description: Causes the EL Load to return the parsed string to the user. The parsed string may not be the com-

plete string sent to the EL; the parsed string contains the short form mnemonic for the command. Fac-
tory default is SYST:COMM:ECHO off.

B.85

SYSTem:COMMunicate:ECHO? QUERY

SYST:COMM:ECHO?

Syntax:

Short Form: SYST:COMM:ECHO?

Long Form: SYSTem:COMMunicate:ECHO?

Returns: 1 if echo is on, 0 if echo is off

Description: Indicates whether echo is on or off.

B.86

SYSTem:COMMunicate:LAN:IPADdress COMMAND

SYST:COMM:LAN:IPAD

Syntax:

Short Form: SYST:COMM:LAN:IPAD
Long Form: SYSTem:COMMunicate:LAN:IPADdress
Where: = numbers from 0 to 255 comprising the IP Address

Description: Sets the IP address to be used during LAN communication.

B.87

SYSTem:COMMunicate:LAN:IPADdress? QUERY

SYST:COMM:LAN:IPAD?

Syntax:

Short Form: SYST:COMM:LAN:IPAD?
Long Form: SYSTem:COMMunicate:LAN:IPADdress?
Returns: = numbers from 0 to 255 comprising the IP Address

Description: Returns the IP Address used during LAN communication.