3 questionable status register, 4 channel status register, Questionable status register -19 – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual

Page 55: Channel status register -19

background image

SERIES EL 071414

3-19

3.13.7.3

QUESTIONABLE STATUS REGISTER

The QUEStionable condition register (see Figure 3-4) contains status bits representing unusual
operation such as inconsistent sensor readings or unusual but non-fault conditions.

0 - Voltage Fault (VF) — 1 indicates the Voltage Fault

1 - Overcurrent (OC) — 1 indicates overcurrent condition has been detected.

2 - Not Used — always zero.

3 - Overpower (OP) — 1 indicates power has exceeded the rated maximum.

4 - Overtemperature (OT) — 1 indicates overtemperature condition has been detected.

5 - 7 - Not Used — always zero.

8 - Not Used — always zero.

9 - Not Used — always zero.

10 - Not Used — always zero.

11 - Not Used — always zero.

12 - Overvoltage (OV) — 1 indicates overvoltage condition has been detected.

13 - Protection Shutdown (PS) — 1 indicates protection shutdown has occurred

14 - 15 - Not Used — always zero.

3.13.7.4

CHANNEL STATUS REGISTER

The Channel Status registers (see Figure 3-4), one for each parallel-connected load, reflect the
state of each load

0 - Voltage Fault (VF) — 1 indicates the Voltage Fault

1 - Overcurrent (OC) — 1 indicates overcurrent condition has been detected.

2 - Not Used — always zero.

3 - Overpower (OP) — 1 indicates power has exceeded the rated maximum.

4 - Overtemperature (OT) — 1 indicates overtemperature condition has been detected.

5 - 7 - Not Used — always zero.

8 - Not Used — always zero.

9 - Not Used — always zero.

10 - Not Used — always zero.

11 - Not Used — always zero.

12 - Overvoltage (OV) — 1 indicates overvoltage condition has been detected.

13 - Protection Shutdown (PS) — 1 indicates protection shutdown has occurred

14 - 15 - Not Used — always zero.