A.7 *opc — operation complete command, A.8 *opc? — operation complete query, A.9 *rst — reset command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual

Page 67

background image

SERIES EL 071414

A-3

A.7

*OPC — OPERATION COMPLETE COMMAND

*OPC

Syntax:

*OPC

Description: Causes load to set status bit 0 (Operation Complete) when pending operations are complete

This command sets Standard Event Status Register bit 0 to “1” when all previous commands have
been executed and changes in output level have been completed. This command does not prevent
processing of subsequent commands, but bit 0 will not be set until all pending operations are com-
pleted. (1 = set = enable function, 0 = reset = disable function).

A.8

*OPC? — OPERATION COMPLETE QUERY

*OPC?

Syntax:

*OPC?
Return value: <1> (ASCII) placed in output queue when load has completed operation.

Description: Indicates when pending operations have been completed.When all pending operations are com-

plete (all previous commands have been executed and changes in output level have been completed)
a “1” is placed in the Output Queue. Subsequent commands are inhibited until the pending operations
are completed. *OPC? is intended to be used at the end of a command line so that the application pro-
gram can monitor the bus for data until it receives the “1” from the Output Queue.

A.9

*RST — RESET COMMAND

*RST

Syntax:

*RST

Description: Causes system reset of EL Load. *RST causes the unit to be reset as follows.

VOLT:PROT:OVE Overvoltage protection set to rated overvoltage protection limit of the
Load.

VOLT:PROT:UND Undervoltage protection set to 0.

CURR:PROT Current Protection set to rated overcurrent protection limit of the load. or
load system when in Master/slave operation.

POW:PROT Power protection set to rated overpower protection of the load or load sys-
tem when in master/slave operation.

CURR Current set to 0.

VOLT Voltage set to maximum voltage of load.

POW Power set to 0.

RES Resistance set to 1000 ohms.

COND Conductance set to .001.

INP Input set to off.

MODE Mode set to current.

SYST:REPLY set to off.

No other features are affected by *RST command.