A.7 *opc — operation complete command, A.8 *opc? — operation complete query, A.9 *rst — reset command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual
Page 67

SERIES EL 071414
A-3
A.7
*OPC — OPERATION COMPLETE COMMAND
*OPC
Syntax:
*OPC
Description: Causes load to set status bit 0 (Operation Complete) when pending operations are complete
This command sets Standard Event Status Register bit 0 to “1” when all previous commands have
been executed and changes in output level have been completed. This command does not prevent
processing of subsequent commands, but bit 0 will not be set until all pending operations are com-
pleted. (1 = set = enable function, 0 = reset = disable function).
A.8
*OPC? — OPERATION COMPLETE QUERY
*OPC?
Syntax:
*OPC?
Return value: <1> (ASCII) placed in output queue when load has completed operation.
Description: Indicates when pending operations have been completed.When all pending operations are com-
plete (all previous commands have been executed and changes in output level have been completed)
a “1” is placed in the Output Queue. Subsequent commands are inhibited until the pending operations
are completed. *OPC? is intended to be used at the end of a command line so that the application pro-
gram can monitor the bus for data until it receives the “1” from the Output Queue.
A.9
*RST — RESET COMMAND
*RST
Syntax:
*RST
Description: Causes system reset of EL Load. *RST causes the unit to be reset as follows.
•
VOLT:PROT:OVE Overvoltage protection set to rated overvoltage protection limit of the
Load.
•
VOLT:PROT:UND Undervoltage protection set to 0.
•
CURR:PROT Current Protection set to rated overcurrent protection limit of the load. or
load system when in Master/slave operation.
•
POW:PROT Power protection set to rated overpower protection of the load or load sys-
tem when in master/slave operation.
•
CURR Current set to 0.
•
VOLT Voltage set to maximum voltage of load.
•
POW Power set to 0.
•
RES Resistance set to 1000 ohms.
•
COND Conductance set to .001.
•
INP Input set to off.
•
MODE Mode set to current.
•
SYST:REPLY set to off.
No other features are affected by *RST command.