beautypg.com

Setting up the oscilloscope, Gds-1000 series user manual – GW Instek GDS-1000 Series User Manual User Manual

Page 18

background image

GDS-1000 Series User Manual

18

Setting up the Oscilloscope

Background

This section describes how to set up the
oscilloscope properly including connecting a
signal, adjusting the scale, and compensating the
probe. Before operating the oscilloscope in a new
environment, run these steps to make sure the
oscilloscope is functionally stable.

Procedure

1. Connect the power cord.

2. Press the power switch. The

display will become active
in approximately 10
seconds.

3. Reset the system by recalling

the factory settings. Press
the Save/Recall key, then
Default Setup. For details of
factory settings, see page34.

4. Connect the probe between the Channel1 input

terminal and probe compensation signal output
(2Vp-p, 1kHz square wave).

5. Set the probe attenuation to x10.

VOLTS/DIV

VOLTS/DIV

TIME/DIV

CH 1

MATH

CH 2

MENU

MENU

Acquire

Display

Utility

Help

Run/Stop

VARIABLE

FORCE

Autoset

Cursor

SINGLE

Hardcopy

Measure Save/Recall

LEVEL

VERTICAL

HORIZONTAL

TRIGGER

Digital Storage Oscilloscope

GDS-1042

40MHz 250M Sa/s

CH1

CAT

300V

MW 22pF

MAX. 300Vpk

1

CH2

EXT TRIG

CAT

300V

MW 22pF

MAX. 300Vpk

1

X

Y

CH1