Common specifications – GW Instek GDS-1000 Series User Manual User Manual
Page 101
APPENDIX
101
Common specifications
Vertical
Sensitivity
2mV/div~5V/Div (1-2-5 increments)
Accuracy
± (3% x |Readout|+0.1div + 1mV)
Bandwidth
See model-specific specifications
Rise Time
See model-specific specifications
Input Coupling
AC, DC, Ground
Input
Impedance
1MΩ±2%, ~15pF
Polarity
Normal, Invert
Maximum Input 300V (DC+AC peak), CAT II
Math Operation +, –, FFT
Offset Range
2mV/div~50mV/div: ±0.4V
10mV/div~500mV/div: ±4V
1V/div~5V/div: ±40V
Trigger
Sources
CH1, CH2, Line, EXT
Modes
Auto, Normal, Single, TV, Edge, Pulse
Coupling
AC, DC, LF rej, HF rej, Noise rej
Sensitivity
See model-specific specifications
External trigger
Range
DC: ±15V, AC: ±2V
Sensitivity
See model-specific specifications
Input Impedance 1MΩ±2%, ~16pF
Maximum Input 300V (DC+AC peak), CATII
Horizontal
Range
1ns/div~10s/div, 1-2.5-5 increment
Roll: 250ms/div – 10s/div
Modes
Main, Window, Window Zoom, Roll, X-Y
Accuracy
±0.01%
Pre-Trigger
10 div maximum
Post-Trigger
1000 div
X-Y Mode
X-Axis Input
Channel 1
Y-Axis Input
Channel 2
Phase Shift
±3° at 100kHz
Signal Acquisition Real-Time
250M Sa/s maximum
Equivalent
25G Sa/s maximum
Vertical
Resolution
8 bits
Record Length
4k points maximum
Acquisition
Normal, Peak Detect, Average
Peak Detection
10ns (500ns/div ~ 10s/div)
Average
2, 4, 8, 16, 32, 64, 128, 256