Output current verification – GW Instek SPD-3606 User Manual

Page 24

background image

PERFORMANCE

VERIFICATION

47

Output Current verification

Check item

Minimum output Current accuracy

Maximum output Current accuracy

Current meter accuracy (Output On)

Current meter accuracy (Output Off)

Connection

Digital Multimeter

SPD-3606

V

A

Verification step

1. Set the SPD panel according to the Default

setting list, page41.

2. Connect SPD Channel 1 and

Digital Multimeter Current
terminal.

CH1

0 60V , 3A

0 30V , 6A

3. Power up SPD and Digital Multimeter.

4. Turn up the SPD Voltage knob to

the maximum.

VOLTAGE

5. Turn On the SPD output.

OUTPUT

ON / OFF

OUTPUT

ON / OFF

SPD-3606 User Manual

48

Minimum output
Current

6. Record the Multimeter reading as the Minimum

output Current. Here is the acceptance range.

Channel1/2

< 0A

7. Turn up the SPD Current knob to

the maximum. Switch the
Multimeter Current terminal to
high current range.

CURRENT

Maximum output
Current

8. Record the Multimeter reading as the Maximum

output Current. Here is the acceptance range.

Channel1/2

6.15A ~ 6.25A

Channel3

3.05A ~ 3.15A

Current meter
accuracy

9. Compare and record the difference between the

SPD Current meter and the Multimeter reading
as the Current meter accuracy (Out On).

Then turn the output Off,
and check the SPD
reading again (Out Off).

OUTPUT

ON / OFF

OUTPUT

ON / OFF

Here is the acceptance range and example.

Channel1/2/3

difference < ±(0.5% of reading +
2 digits) of Multimeter

Example:
Multimeter (Out On) = 3.000A
Tolerance = ±(0.005*3 + 0.02) ≈ ±0.04A
Accepted SPD reading(Out On)= 2.96A~3.04A
Accepted SPD reading (Out Off)= 2.96A~3.04A

10. Connect the Multimeter to

SPD Channel2 and repeat step
4 to 9.

CH2

0 60V , 3A

0 30V , 6A