OPERATION

31

2. Set the tracking switch

position to Series, + .

TRAC

KING

INDE

P.

SERI

ES

PARA

LLEL

3. Set the OVP if necessary. In

tracking series mode, set the
Channel2 (Slave) OVP setting to
the maximum level, so that the
OVP trips if the Channel1
(Master) setting is violated. For
OVP setup details, see page22.

O.V.P.

SET

TRIP

4. Connect the load to the front panel terminals,

channel1+ & channel2− (Single supply).

CH2

CH1

GND

0 60

V , 3A

0 30

V , 6A

0 60

V , 3A

0 30

V , 6A

5. Turn up the Channel2

Current knob to maximum.

CURRENT

CH2

SLAVE

6. Set the output Voltage and

Current using the Channel1
(Master) knob.

VOLTAGE

CURRENT

CH1

MASTER

C.V.

C.C.

SPD-3606 User Manual

32

7. Refer to the Channel1 (Master) meter and

indicator for the output setting level and
CV/CC status.

CH1 / CH3

CH1

MASTER

C.V.

C.C.

Current

level

Channel1 meter reading shows
the output Current. (Channel2
Current control must be in the
Maximum position).

Voltage

level

Double the reading on the
Channel1 Voltage meter. (In the
above case, the actual output is
23.6 x 2 = 47.2V).

8. Press the Output key. The

Output indicator turns
green.

OUTPUT

ON / OFF

OUTPUT

ON / OFF