Output voltage verification, Default spd-3606 setting – GW Instek SPD-3606 User Manual

Page 21

background image

PERFORMANCE

VERIFICATION

41

Default SPD-3606 setting

The following is the required front and rear panel setting before
running each verification.

Range

60V/3A

60V/3A

30V/6A

30V/6A

60V/3A

30V/6A

Output Voltage

Tracking Series

OVP

Output Current

Tracking Parallel

OVP setting SW

Normal position

CH3

CH2

CH1

O.V.P. SET

NORMAL

Channel1/3 meter
SW

CH1 position

CH1

CH3

Tracking SW

Independent position, + .

TRAC

KING

INDE

P.

SERI

ES

PARA

LLEL

Channel1/2/3
Voltage knob

Minimum position

VOLTAGE

Channel1/2
Current knob

Minimum position

CURRENT

SPD-3606 User Manual

42

Output Voltage Verification

Check item

Minimum output Voltage accuracy

Maximum output Voltage accuracy

Voltage meter accuracy (Output On)

Voltage meter accuracy (Output Off)

Connection

Digital Multimeter

SPD-3606

V

A

Verification step

1. Set the SPD panel according to the Default

setting list, page41.

2. Connect SPD Channel 1 and

Digital Multimeter Voltage
terminal.

CH1

0 60V , 3A

0 30V , 6A

3. Power up SPD and Digital Multimeter.

4. Turn up the SPD Current knob to

the maximum.

CURRENT