Emote output control, Erformance verification – GW Instek SPD-3606 User Manual
Page 20
REMOTE OUTPUT CONTROL
39
R
EMOTE OUTPUT
CONTROL
Background
The remote control terminal on the
rear panel can turn the output
On/Off, just like the Output key on
the front panel. This feature is useful
for automated measurement and
testing using externally connected
control device, such as in production
line or incoming quality inspection.
OUTPUT
ON / OFF
Wire connection
Connect the terminals using
bare wires. Use a minus
driver to push the orange
part, insert the wire, then
release the orange part.
1
2
3
Remote control
Off (front panel
control)
When there is no connection,
Output On/Off is entirely
controlled from the front panel.
Output always Off
When the left two terminals are
connected, the Output is always
Off. The front panel Output key is
disabled.
Output always On
When the right two terminals are
connected, the Output is always
On. The front panel Output key is
disabled.
SPD-3606 User Manual
40
P
ERFORMANCE
VERIFICATION
Overview
Background
Performance verification checks SPD functionality
before the operation or at the incoming inspection.
Recording tables are attached at the end of this
chapter.
Verification item
•
Output Voltage
•
Tracking Series Voltage
•
Output Current
•
OVP
Equipment
Digital Multimeter
•
DCV Accuracy < 0.1%
•
DCA Accuracy < 0.5%
•
DCA range: ≥ 12A
•
Resolution ≥ 4 ½ digit
•
Recommended model: GDM-8245, GDM-8246
Multimeter
– SPD cable
•
Voltage rating > 70V
•
Current rating > 12A
Philips screw
driver
•
< 3mm (for OVP adjustment)