Ch1/ch2 tracking parallel mode – GW Instek SPD-3606 User Manual

Page 18

background image

OPERATION

35

6. Set the output Current

separately, using both the
Channel1 (Master) and
Channel2 (Slave) Current
knob.

VOLTAGE

CURRENT

CH1

MASTER

C.V.

C.C.

VOLTAGE

CURRENT

CH2

SLAVE

C.V.

C.C.
PAR.

CH2

CH1 / CH3

CH1(+)~COM Current = 1.84A in the above case

CH2(–)~COM Current = 2.18A in the above case

7. Press the Output key. The

Output indicator turns
green.

OUTPUT

ON / OFF

OUTPUT

ON / OFF

8. Refer to the Channel1

(Master) indicator for
CH1(+) ~ COM CV/CC
status, and the Channel2
(Slave) indicator for
CH2(–) ~ COM CV/CC
status.

CH1

MASTER

C.V.

C.C.

CH2

SLAVE

C.V.

C.C.
PAR.

SPD-3606 User Manual

36

CH1/CH2 Tracking Parallel Mode

Background /
Connection

Tracking parallel operation doubles the Current
capacity of SPD-3606 by internally connecting
Channel1 (Master) and Channel2 (Slave) in parallel
and combining the output to a single channel.
Channel1 (Master) controls the combined output.

CH2

CH1

LOAD

Output rating

0 ~ 30V/0 ~ 12A or 0 ~ 60V/0 ~ 6A

Setting step

1. Select the output range,

60V/3A(6A) or
30V/6A(12A). Set the
CH1/CH3 meter switch to
the CH1 position.

60V/3A

30V/6A

CH1

CH3

2. Set the tracking switch

position to Parallel, + .

TRAC

KING

INDE

P.

SERI

ES

PARA

LLEL