Ch1/ch2 tracking parallel mode – GW Instek SPD-3606 User Manual
Page 18

OPERATION
35
6. Set the output Current
separately, using both the 
Channel1 (Master) and 
Channel2 (Slave) Current 
knob. 
VOLTAGE
CURRENT
CH1
MASTER
C.V.
C.C.
VOLTAGE
CURRENT
CH2
SLAVE
C.V.
C.C.
PAR.
CH2
CH1 / CH3
CH1(+)~COM Current = 1.84A in the above case
CH2(–)~COM Current = 2.18A in the above case
7. Press the Output key. The
Output indicator turns 
green. 
OUTPUT
ON / OFF
OUTPUT
ON / OFF
8. Refer to the Channel1
(Master) indicator for 
CH1(+) ~ COM CV/CC 
status, and the Channel2 
(Slave) indicator for 
CH2(–) ~ COM CV/CC 
status. 
CH1
MASTER
C.V.
C.C.
CH2
SLAVE
C.V.
C.C.
PAR.
SPD-3606 User Manual
36
CH1/CH2 Tracking Parallel Mode
Background / 
Connection 
Tracking parallel operation doubles the Current 
capacity of SPD-3606 by internally connecting 
Channel1 (Master) and Channel2 (Slave) in parallel 
and combining the output to a single channel. 
Channel1 (Master) controls the combined output. 
CH2
CH1
LOAD
Output rating
0 ~ 30V/0 ~ 12A or 0 ~ 60V/0 ~ 6A
Setting step
1. Select the output range,
60V/3A(6A) or 
30V/6A(12A). Set the 
CH1/CH3 meter switch to 
the CH1 position. 
60V/3A
30V/6A
CH1
CH3
2. Set the tracking switch
position to Parallel, + .
TRAC
KING
INDE
P.
SERI
ES
PARA
LLEL