PERFORMANCE

VERIFICATION

43

5. Turn On the SPD output.

OUTPUT

ON / OFF

OUTPUT

ON / OFF

Minimum output
Voltage

6. Record the Multimeter reading as the Minimum

output Voltage. Here is the acceptance range.

Channel1/2

< 0V

Channel3

< 100mV

7. Turn up the SPD Voltage knob to

the maximum. Switch the
Multimeter Voltage terminal if
necessary.

VOLTAGE

Maximum output
Voltage

8. Record the Multimeter reading as the Maximum

output Voltage. Here is the acceptance range.

Channel1/2

61.5V ~ 62.5V

Channel3

5.20V ~ 5.30V

Voltage meter
accuracy

9. Compare and record the difference between the

SPD Voltage meter and the Multimeter reading
as the Voltage meter accuracy (Out On).

Then turn the output Off,
and check the SPD
reading again (Out Off).

OUTPUT

ON / OFF

OUTPUT

ON / OFF

Here is the acceptance range and example.

Channel1/2/3

difference < ±(0.5% of reading +
2 digits) of Multimeter

SPD-3606 User Manual

44

Example:
Multimeter (Out On) = 30.00V
Tolerance = ±(0.005*30 + 0.2) ≈ ±0.4V
Accepted SPD reading(Out On)= 29.6V~30.4V
Accepted SPD reading (Out Off)=29.6V~30.4V

Channel2

10. Connect the Multimeter to

SPD Channel2 and repeat step
4 to 9.

CH2

0 60V , 3A

0 30V , 6A

Channel3

11. Switch the CH1/CH3

meter switch to CH3
position. Connect the
Multimeter to Channel3
and repeat step 5 to 9.

CH1

CH3

CH3

0.1 5V , 3A

* Skip step 4 since Channel 3 does not have

Current knob.