PERFORMANCE
VERIFICATION
43
5. Turn On the SPD output.
OUTPUT
ON / OFF
OUTPUT
ON / OFF
Minimum output 
Voltage 
6. Record the Multimeter reading as the Minimum
output Voltage. Here is the acceptance range.
Channel1/2
< 0V
Channel3
< 100mV
7. Turn up the SPD Voltage knob to
the maximum. Switch the 
Multimeter Voltage terminal if 
necessary. 
VOLTAGE
Maximum output 
Voltage 
8. Record the Multimeter reading as the Maximum
output Voltage. Here is the acceptance range.
Channel1/2
61.5V ~ 62.5V
Channel3
5.20V ~ 5.30V
Voltage meter 
accuracy 
9. Compare and record the difference between the
SPD Voltage meter and the Multimeter reading 
as the Voltage meter accuracy (Out On). 
Then turn the output Off, 
and check the SPD 
reading again (Out Off). 
OUTPUT
ON / OFF
OUTPUT
ON / OFF
Here is the acceptance range and example.
Channel1/2/3
difference < ±(0.5% of reading + 
2 digits) of Multimeter 
SPD-3606 User Manual
44
Example: 
Multimeter (Out On) = 30.00V 
Tolerance = ±(0.005*30 + 0.2) ≈ ±0.4V 
Accepted SPD reading(Out On)= 29.6V~30.4V 
Accepted SPD reading (Out Off)=29.6V~30.4V 
Channel2
10. Connect the Multimeter to
SPD Channel2 and repeat step 
4 to 9. 
CH2
0 60V , 3A
0 30V , 6A
Channel3
11. Switch the CH1/CH3
meter switch to CH3 
position. Connect the 
Multimeter to Channel3 
and repeat step 5 to 9. 
CH1
CH3
CH3
0.1 5V , 3A
* Skip step 4 since Channel 3 does not have
Current knob.