LOGIC ANALYZER

69

Trigger Settings

Serial Bus Trigger Settings

UART BUS Trigger Settings

The UART bus trigger conditions can be set at any time after the
bus settings have been set to UART.

Panel Operation

1. Set the Bus to UART in the bus

menu.

Page 54

2. Press the Trigger Menu key.

Menu

3. Press Type from the bottom menu.

4. Press Others from the side

menu and select Bus.

5. Press Trigger On and select the

triggering condition for the UART
bus.

Trigger On

Tx Start Bit, Rx Start Bit, Tx End of
Packet, Rx End of Packet, Tx Data,
Rx Data, Tx Parity Error, Rx Parity
Error

Trigger On – Tx
Data, Rx Data

If Tx Data or Rx Data was configured for the
Trigger On setting, then the number of bytes and
data can also be configured.

1. Press Data from the bottom menu.