beautypg.com

Function generator operation, Overview, Function generator specifications – GW Instek GDS-2000A series Option User Manual User Manual

Page 20

background image

GDS-2000A Series Options Manual

20

Function Generator Operation

Overview

Background

The DDS Function Generator module allows the
GDS-2000A to create basic sine, square and
triangle waveforms.

Note

The function generator option can only be used
with firmware version 1.13 or later. However it is
highly recommended that firmware version V1.16
or later be installed to access the full functionality
of the module. The instructions on the following
pages are based on V1.16 or later.

Please see the GW Instek website for the latest
firmware and the firmware installation procedure.

Function Generator Specifications

Waveforms

Sine, Square, Triangle

Frequency range

0.1Hz ~ 5MHz for sine
0.1Hz ~ 5MHz for square
0.1Hz ~ 500KHz for triangle

Frequency Stability

±50ppm

Frequency Accuracy

±50ppm (± 0.25Hz)

Aging

±5ppm/Year

Amplitude Range

60mVpp ~ 6Vpp (into 1MΩ)
30mVpp ~ 3Vpp (into 50Ω)

Amplitude Accuracy ±10%
Attenuator

-20dB

Impedance

50Ω

DC Offset

±2V (into 1MΩ)
±1V (into 50Ω)

Duty Control Range

5%(min) ~ 95%(max)
(Square wave only)

Rise or Fall Time

15ns (Square wave only)