20
Function Generator Operation
Overview
Background
The DDS Function Generator module allows the
GDS-2000A to create basic sine, square and
triangle waveforms.
Note
The function generator option can only be used
with firmware version 1.13 or later. However it is
highly recommended that firmware version V1.16
or later be installed to access the full functionality
of the module. The instructions on the following
pages are based on V1.16 or later.
Please see the GW Instek website for the latest
firmware and the firmware installation procedure.
Function Generator Specifications
Waveforms
Sine, Square, Triangle
Frequency range
0.1Hz ~ 5MHz for sine
0.1Hz ~ 5MHz for square
0.1Hz ~ 500KHz for triangle
Frequency Stability
±50ppm
Frequency Accuracy
±50ppm (± 0.25Hz)
Aging
±5ppm/Year
Amplitude Range
60mVpp ~ 6Vpp (into 1MΩ)
30mVpp ~ 3Vpp (into 50Ω)
Amplitude Accuracy ±10%
Attenuator
-20dB
Impedance
50Ω
DC Offset
±2V (into 1MΩ)
±1V (into 50Ω)
Duty Control Range
5%(min) ~ 95%(max)
(Square wave only)
Rise or Fall Time
≤
15ns (Square wave only)