Function Generator Calibration

Background

The function generator module can also be
calibrated from the System menu. The calibration
function is only available for firmware version
V1.16 or later.

Connection

Connect the GEN 1 or GEN 2 output (depending
on which module slot the function generator is
installed in) to CH1 using a BNC cable.

CH1

GEN 1

or

GEN 2

Note

If two function generator modules are installed,
the system will automatically choose to calibrate
the function generator in module slot 1. If you
wish to calibrate the second function generator,
first select it from the Option menu (Option key >
Function Gen 2).

Panel Operation

1. Press the Utility key.

Utility

2. Press System from the bottom

menu.

3. Press More 1 of 2 from the side

menu.

4. Press Self CAL from the side menu.