<8. Detailed Data Setting>

8-5

IM 11M12A01-04E

8.3.2 Preference Order of Output Hold Value

The output hold value takes the following preference order:

During error occurrence
During calibration
During maintenance
During warm-up

Preference order (high)

8.3.2E.siki

For example, if the output current is set to “4 mA” during maintenance, and “without hold” output

during calibration is preset, the output is held at 4 mA in the maintenance display. However, the output

hold is released at the time of starting the calibration, and the output will be held again at 4 mA after

completing the calibration and when the output stabilization time elapses.

8.3.3 Output Hold Setting

Table 8.5 lists parameter codes with set values for individual set items.

Table 8.5

Parameter Codes for Output Holding

Set items

Parameter code

Set value

During warm-up

C04

0 4 mA
1 20 mA
2 Holds Set value

During maintenance

C05

0 Without hold feature
1 Last measured value.
2 Holds set values.

During calibration

C06

0 Without hold feature
1 Last measured value.
2 Holds set values.

During error occurrence

C07

0 Without hold feature
1 Last measured value.
2 Holds set values.

Note: "C07" is not displayed when option code "/C2" or "/C3" (NAMUR NE 43 compliant) is specified.

8.3.4 Default Values

When the analyzer is delivered, or if data are initialized, output holding is by default as shown in Table

8.6.

Table 8.6

Output Hold Default Values

Status

Output hold (min. and max. values)

Preset value

During warm-up

4 mA

4 mA

Under maintenance

Holds output at value just before maintenance started.

4 mA

Under calibration or blow-back Holds output at value just before starting calibration

4 mA

On Error occurrence

Holds output at a preset value.

3.4 mA