<2. Specifications>

2-12

IM 11M12A01-04E

2.2.2 ZR20H Automatic Calibration Unit

This automatic calibration unit is applied to supply specified flow of reference gas and calibration gas

during automatic calibration to the detector in a system configuration (System 3).

Specifications

Equipped with the analyzer when automatic calibration is specified in the suffix code of the ZR202G

Integrated type by selecting either “-A (Horizontal mounting)” or “-B (Vertical mounting)”. The ZR20H

should be arranged when automatic calibration is to be required after the ZR202G has been installed.

Ask Yokogawa service station for its mounting.
Construction:

Dust-proof and rainproof construction; NEMA4X/IP67 (excluding flowmeter)

Mounting:

Mounted on ZR202G, no vibration

Materials:

Body; Aluminum alloy, Piping; SUS316 (JIS), SUS304 (JIS), Flowmeter;

MA (Methacrylate resin), Bracket; SUS304 (JIS)

Finish:

Polyurethane corrosion-resistance coating

Case; Mint green (Munsell 5.6 BG3.3 /2.9)

Cover; Mint green (Munsell 5.6 BG3.3/2.9)

Piping Connection:

Rc 1/4 or 1/4 NPT

Power Supply:

24V DC (from ZR202G)

Power consumption:

Approx.1.3 W

Reference Gas Pressure: Sample gas pressure plus Approx. 150 kPa (690 kPa max.), (Pressure at

inlet of automatic calibration unit)

Air Consumption:

Approx. 1.5 l/min

Weight:

Approx. 2 kg

Ambient Temperature: -20 to +55°C, no condensing and freezing
Ambient Humidity:

0 to 95% RH

Storage Temperature: -30 to +65°C

• Model and Codes

Model

Suffix code

Option code

Description

ZR20H

- - - - - - - - - - - - - - - - - - - - - - - - - - - - - Automatic calibration unit for ZR202G *1

Gas piping

connection

-R

-T

- - - - - - - - - - - -

- - - - - - - - - - - -

Rc 1/4

1/4 NPT

Reference gas *2

-E

-P

- - - - - - - - - - - -

- - - - - - - - - - - -

Instrument air

Pressure compensated

Mounting

-A

-B

- - - - - - - - - - - -

- - - - - - - - - - - -

Horizontal mounting

Vertical mounting

-A

- - - - - - - - - - - - Always -A

*1 Ask Yokogawa service station for additional mounting of ZR20H to the preinstalled ZR202G.