3 initialization procedure, Initialization procedure -8, F3 0 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 117: Usr go

<10. Other Functions>
10-8
IM 11M12A01-04E
10.3 Initialization Procedure
Follow the table below to initialize parameters. The password for initialization is 1255.
Table 10.5
Initialization Procedure
Switch operation
Display
Description
>
∧
ENT
F
3
0
Enter the parameter code for the item to be initialized.
The following show an example of entering "F30."
(Previous needed operations are omitted.)
>
∧
ENT
0
000
Touch the [ENT] key to switch to the password entry display.
>
∧
ENT
1
000
Enter the password 1255 for initialization.
>
∧
ENT
1
0
00
>
∧
ENT
1
2
00
>
∧
ENT
12
0
0
>
∧
ENT
12
5
0
>
∧
ENT
125
0
>
∧
ENT
125
5
>
∧
ENT
1255
After you enter the password and then touch the [ENT] key, all the digits
flash.
>
∧
ENT
USr Go
Touch the [ENT] key again to display "USr Go."
>
∧
ENT
USr Go
Touch the [ENT] key once more. All the digits then flash for two to three
seconds, and data initialization starts.
>
∧
ENT
F
30
The initialization is complete, and the parameter code selection display
then appears.
The symbol (
) indicates that the keys are being touched, the light characters indicates flashing.
WARNING
• Do not attempt to turn off the equipment power during initialization (while “USr Go” is flashing).