Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 123

<10. Other Functions>
10-14
IM 11M12A01-04E
Contact-related Items in Group E
Code
Item
Tuning
Engineering unit
Default setting
E01 Selection of input contact 1 0 Invalid
Invalid
1 Calibration gas pressure decrease
2 Measurement range change
3 Calibration start
4 Detection of non-combusted gas
E02 Selection of input contact 2 0 Invalid
Invalid
1 Calibration gas pressure decrease
2 Measurement range change
3 Calibration start
4 Detection of non-combusted gas
E03 Selecting action of input
contact 1
0 Action with closed contact
Action with closed
contact
1 Action with open contact
E04 Selecting action of input
contact 2
0 Action with closed contact
Action with closed
contact
1 Action with open contact
E10 Selecting action of output
contact 1
0 Action with closed contact
(normally energized)
Action with closed
contact
1 Action with open contact (normally
de-energized)
E20 Output contact 1 error
0 No action
No action
1 Action
E21 Output contact 1, high-high
alarm
0 No action
No action
1 Action
E22 Output contact 1, high
alarm
0 No action
No action
1 Action
E23 Output contact 1, low alarm 0 No action
No action
1 Action
E24 Output contact 1, low-low
alarm
0 No action
No action
1 Action
E25 Output contact 1, during
maintenance
0 No action
Action
1 Action
E26 Output contact 1, during
calibration
0 No action
No action
1 Action
E27 Output contact 1,
measurement range
change
0 No action
No action
1 Action
E28 Output contact 1, during
warm up
0 No action
Action
1 Action
E29 Output contact 1,
calibration gas
pressure decrease
0 No action
No action
1 Action
E32 Output contact 1, detection
of non- combusted gas
0 No action
No action
1 Action