beautypg.com

Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 107

background image

<9. Calibration>

9-6

IIM 11M12A01-04E

9.2.1.4

Span gas Concentration

Set the oxygen concentration for span calibration. If instrument air is used as the span gas, enter 21

%O

2

.

When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas), use a hand-

held oxygen analyzer to measure the actual oxygen concentration, and then enter it.

CAUTION

(1) When instrument air is used for the span calibration, remove the moisture from the instrument air

at a dew-point temperature of -20°C and also remove any oily mist and dust from that air.
(2) If dehumidifying is not enough, or if foul air is used, the measurement accuracy will be adversely

affected.

9.2.1.5

Calibration Time

• When the calibration mode is in manual:

First set the hold (output stabilization) time. This indicates the time required from the end of

calibration to entering a measurement again. This time, after calibration, the measurement

gas enters the sensor to set the time until the output returns to normal. The output remains

held after completing the calibration operation until the hold (output stabilization) time elapses.

The calibration time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. For

more details, consult Section 8.3,”Setting Output Hold.” When the calibration mode is in semi-

automatic, set the hold (output stabilization) time and calibration time. The calibration time is the

time required from starting the flow of the calibration gas to reading out the measured value. The

set calibration time is effective in conducting both zero and span calibrations. The calibration

time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. Figure 9.4 shows the

relationship between the calibration time and hold (output stabilization) time.

Analog output remains hold

(when output remains hold)

Calibration start

(contact or switch input)

Span calibration

(span gas valve open)

Zero calibration

(zero gas valve open)

Analog output status

Calibration time

Calibration time

Hold (output stabilization)

time

F9.4E.ai

Figure 9.4

Calibration and Hold (Output stabilization) Time Settings

• When the calibration mode is in automatic:

In addition to the above hold (output stabilization) time and calibration time, set the interval, start

date, and start time.

Interval means the calibration intervals ranging from 000 days, 00 hours to 255 days, 23 hours.

Set the first calibration day and the start-calibration time to the start date and start time

respectively. After the first calibration is carried out, the next calibration will be executed

according to the preset calibration intervals.