2 wiring procedure, Wiring procedure -6 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 63
<5. Wiring>
5-6
IM 11M12A01-04E
5.5.2 Wiring Procedure
(1) M4 screws are used for the terminal of the converter. Each cable should be equipped with the
corresponding crimp contact.
(2) The ON/OFF level of this contact input is identified by the resistance. Connect a contact input
that satisfies the descriptions in Table 5.2.
Table 5.2
Identification of Contact Input ON/OFF
Closed
Open
Resistance
200 Ω or less
100 kΩ or more