<8. Detailed Data Setting>
8-3
IM 11M12A01-04E
8.3 Output Hold Setting
The “output hold” functions retain an analog output signal at a preset value during the equipment’s
warm-up time or calibration or if an error arises.
Table 8.4 shows the analog outputs that can be retained and the individual states.
Table 8.4
Analog Output Hold Setting
Equipment status
Output hold
values available
During warm-up During
maintenance
During calibration Error occurrence
(*1)
4 mA
O
20 mA
O
Without hold feature
O
O
O
Retains output from just
before occurrence
O
O
O
Set value (2.4 to 21.6
mA)
O
O
O
O
O:
The output hold functions are available.
*1: The output hold functions on error occurrence are unavailable when option code "/C2" or "/C3" (NAMER NE 43 compliant) is
specified.
8.3.1 Definition of Equipment Status
(1) During warm-up
“During warm-up” is the time required after applying power until the sensor temperature
stabilizes at 750°C, and the equipment is in the measurement mode. This status is that the
sensor temperature is displayed on the basic panel.
(2) During maintenance
“During maintenance” is the time from when a valid password is entered in the basic panel
display to enable the parameter code selection display until the display goes back to the basic
panel display
(3) During calibration (see Chapter 9, Calibration)
In the manual calibration, proceed with the calibration operation with the parameter code
「B10」 to display the span gas confirmation display for the first span calibration, thus starting the
calibration time when the [ENT] key is touched. After a series of calibrations is complete and the
preset output stabilization time has elapsed, the calibration time will be up. Figure 8.1 shows the
definition of “during calibration” in the manual calibration.