3.2.2 Mounting of ZA8F Flow Setting Unit

The flow setting unit can be mounted either on a pipe (nominal JIS 50 A) or on a wall. It should be

positioned vertically so that the flowmeter works correctly.

(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting the

flow setting unit. (The unit weighs approximately 2 to 3.5 kg.)

(2) Mount the flow setting unit on the pipe by tightening the nuts with the U-bolt so that the metal

fitting is firmly attached to the pipe.

F3401E.ai

Figure 3.12

Pipe Mounting

(1) Make a hole in the wall as illustrated in Figure 3.13.

223

140

F3.13E.ai

4 - Ø6 hole, or M5 screw

Unit: mm

Figure 3.13

Mounting holes