beautypg.com

Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 136

background image

<11. Inspection and Maintenance>

11-7

IM 11M12A01-04E

CAUTION

When operating an instrument such as boiler or industrial furnace is stopped with the zirconia oxygen

analyzer operation, moisture can condensate on the sensor portion and dusts may stick to it.
If operation is restarted in this condition, the sensor which is heated up to 750°C firmly fixes the dusts

on itself. Consequently, the dusts can make the sensor performance very lower. If a large amount of

water is condensed, the sensor can be broken and never be used.
To prevent the above nonconformity, take the following action when stopping operation.

(1) If possible, keep on supplying the power to converter and flowing reference gas to the sensor.

If impossible to do the above, remove the detector.

(2) If unavoidably impossible to supply the power and removing the detector, keep on following air at

600 ml/min into the calibration gas pipe.

When restarting operation, be sure to flow air, for 5-10 minutes, at 600 ml/min into the calibration gas

pipe before supplying the power to converter.