<10. Other Functions>

10-20

IM 11M12A01-04E

Table 10.9

Sample gas pressure, (kPa)

50

100

150

200

250

Flow rate, (ml/min)

500

430

380

350

320

(3) Adjust the flow rate. After the measured oxygen concentration is stabilized, touch the [ENT] key,

then all the digits will flash. Touch the [ENT] key again to flash “CAL End.”

(4) Close the zero gas flow setting valve to stop the zero gas flow. Be sure to tighten the lock

nut if the valve shaft has a lock nut to prevent any leakage of zero gas into the sensor during

measurement. When the stabilization time elapses, the zero calibration will be complete.

10.6.4 Treatment After Calibration

No special treatment of the instrument is needed after calibration. However, it is recommended that

the pressure regulator for the zero gas cylinders be closed because calibration is not required so

often.