ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 62

R Series Service Manual
2-10
9650-0903-01 Rev. J
POOR PAD CONTACT
Electrode impedance exceeds threshold.
•
Ensure pads are coupled to patient.
•
Check/replace pads or universal cable.
•
Check impedance circuit calibration.
•
Replace system board.
PRESS ANALYZE
Alternate message for Check Patient message if this
configuration option is selected.
PRESS CHARGE
•
Discharge button is pressed in Manual Defib mode but
the unit is not charged.
•
Advisory message following SHOCK ADVISED in
Manual/Advisory Defib mode with auto-charge
disabled.
PRESS SHOCK
Prompt issued in AED auto defib mode when defib is
charged (ready).
READING LOG DATA
Summary Report Log data printing message.
RECORDER FAULT 142
Strip chart system error.
Replace printer assembly, digital board, and/or
printer interconnect board.
RECORDER FAULT 143
Strip chart failed power-up echo test.
Replace printer assembly, digital board, and/or
printer interconnect board.
RECORDER FAULT 147
Strip chart printhead over safe operating temperature.
Replace printer assembly, digital board, and/or
printer interconnect board.
RELEASE BUTTONS
Simultaneous external paddle button presses detected
before unit reached full defib charge (ready state).
Release buttons.
RELEASE SHOCK
•
Discharge switch(es) closed when pressing charge
button.
•
Discharge button pressed before defib reached ready
state.
•
Release shock button.
•
Check paddles.
•
Replace controls board.
REMOVE SYNC
Analyze button pressed, or no heart rate detected while in
Defib Sync mode.
Error Message
Explanation
User
Advisory
Technical Action