Maintenance tests, Overview, Chapter 1 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 11: Overview -1, Chapter 1 maintenance tests

R Series Service Manual
9650-0903-01 Rev. J
1-
1
Chapter 1
Maintenance Tests
Overview
The R Series has two checkout procedures: the R Series Operator’s Guide defibrillator testing checklist and the extensive 12-month maintenance test
checkout procedures.
Because the R Series units must be maintained ready for immediate use, regular readiness testing is required. It can either be performed manually or
automatically. Refer to the R Series Operator’s Guide for details.
A qualified biomedical technician must perform a more thorough maintenance test checkout every 12 months to ensure that the functions of the R Series
unit work properly. This chapter describes the step by step procedures for performing the 12 month maintenance test checkout. Use the checklist at the
back of this document (ZOLL R Series Maintenance Test Checklist) to record your results of the maintenance tests.
This chapter describes the following maintenance tests:
•
1.0 Physical Inspection of the Unit
•
•
•
4.0 Power Supply Test (Optional)
•
•
•
•
8.0 Calibrating Pulses on Strip Chart Test
•
•
•
•
12.0 Synchronized Cardioversion Test
•
13.0 Synchronized Cardioversion for Remote ECG Monitoring Test (Optional)
•
•