Defib 120j sel, 2 charge defibrillator – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 415

Emergency Defibrillation Procedure with Hands-Free Therapy Electrodes
9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
3–9
The selected energy level is shown as DEFIB XXXJ SEL. on the display.
If you have configured Shocks 1, 2, and 3 to escalating energy levels (see the R Series
Configuration Guide for instructions), the R Series automatically sets the energy to the
preconfigured Energy Level: Shock 1, 2, 3 setting at power-up and after each of the first two
shocks. The ENERGY INCREMENTED message will be displayed after Shocks 1 and 2 are
delivered. Manually changing the energy level outside the preprogrammed sequence and
delivering a shock disables this function.
2 Charge Defibrillator
Press the CHARGE button on the front panel.
To increase or decrease the selected energy after you have pressed the CHARGE button, use
the defibrillator ENERGY SELECT buttons.
Caution
Changing the selected energy while the unit is charging or charged causes the defibrillator to
disarm itself. Press the CHARGE button again to charge the unit.
After charging to the selected energy, the SHOCK button on the front panel lights. A
distinctive charge ready tone sounds and the DEFIB XXXJ READY is displayed. The
defibrillator is now ready to discharge.
DEFIB 120J SEL.
Code
Marker
Options