R Series Service Manual
9650-0903-01 Rev. J
2-7
ENERGY INCREMENTED
Defib energy has been automatically incremented to the
next configured level after shock 1 or 2 has been delivered
and the unit is configured for Basic Energy Auto
Escalation.
ERASING REPORT
The unit is erasing the selected report data.
FULL DISCLS STOPPED
Full disclosure data exceeds the storage capacity: 4 hours
or 32Mb.
IF NO PULSE
Message displayed upon entered CPR period specified by
AHA protocol.
INSERT CARD
No card detected during manual or semi-automatic modes.
Re-insert CF Card or replace communication
module.
INT. DUMP OVERLOAD
More than 15 internal discharges in 5 minutes.
Unit needs to cool down, wait approximately
5 minutes.
INVALID GAS TEMP:
RETRY
The operator has attempted to set a gas temperature
outside the Capnostat’s operating range.
Recalibrate EtCO
2
, verify range.
LOW BATTERY
Low battery.
•
Replace battery pack with a fully charged
battery pack.
•
Plug unit into AC mains.
NIBP ARTIFACT
The unit is unable to detect systolic, diastolic or mean
blood pressure due to excessive motion or vibration.
•
Take a single blood pressure measurement.
•
Keep patient as still as possible.
•
Insulate patient, cuff and hose from vibrations
as much as possible.
NIBP COMM ERR 259
Framing, parity, or fifo error.
Replace parameter power supply or NIBP module.
NIBP COMM ERR 260
Received invalid packets.
Replace parameter power supply or NIBP module.
NIBP COMM ERR 261
Device not taking action request.
Replace parameter power supply or NIBP module.
NIBP COMM ERR 262
No reply from device.
Replace parameter power supply or NIBP module.
Error Message
Explanation
User
Advisory
Technical Action