0 paddles test, Paddles test -15 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 25

R Series Service Manual
9650-0903-01 Rev. J
1-
15
6.0 Paddles Test
Tools Needed: None.
Test Setup: If applicable, connect the OneStep cable to the paddles and place the paddles in paddle wells.
Do this...
Observe this...
Pass/Fail/NA
6.1
Turn the selector switch to DEFIB (for AED 
units, turn the selector switch to ON, and 
select Manual Mode.) Press and hold the 
ENERGY DOWN button on the sternum paddle.
The energy selection decreases to 1J.
o
o
o
6.2
Press and release the ENERGY UP button on 
the sternum paddle for each setting.
Verify the energy levels increment through the following: 1-10, 15, 20, 30, 50, 
75, 100, 120, 150, 200J.
o
o
o
6.3
Press and release the RECORDER button on 
the sternum paddle.
The recorder turns on.
o
o
o
6.4
Press and release the RECORDER button 
again.
Verify the recorder turns off.
o
o
o
6.5
Select 30J using the paddle ENERGY button. 
Press the CHARGE button on the Apex paddle.
The unit charges to 30J, then the red LED charge indicator illuminates and the 
charge tone sounds. (Note that the front panel shock button does not 
illuminate).
o
o
o
6.6
Press and release the APEX SHOCK button.
No discharge.
o
o
o
6.7
Press and release the STERNUM SHOCK 
button.
No discharge.
o
o
o
6.8
Press and hold both paddles SHOCK buttons.
The unit discharges. The 30J TEST OK message displays and the red LED 
turns off. The recorder runs. 
o
o
o
Record your results on the Maintenance Test Checklist.