ZOLL R Series Monitor Defibrillator Rev J User Manual
R series, Service manual
Table of contents
Document Outline
- © 2013 ZOLL Medical Corporation. All rights reserved. R Series, CodeNet, SurePower, OneStep and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries. Masimo is a registered trademark of...
- Software License
- Preface
- Maintenance Tests
- 1.0 Physical Inspection of the Unit
- 2.0 Front Panel Button Test
- 3.0 3 and 5 Leads Test
- 4.0 Power Supply Test (Optional)
- 5.0 Leakage Current Test
- 6.0 Paddles Test
- 7.0 Heart Rate Display Test
- 8.0 Calibrating Pulses on Strip Chart Test
- 9.0 Notch Filter Test
- 10.0 Heart Rate Alarm Test
- 11.0 Defibrillator Self Test
- 12.0 Synchronized Cardioversion Test
- 13.0 Synchronized Cardioversion for Remote ECG Monitoring Test (Optional)
- 14.0 Shock Test
- 15.0 Summary Report Test
- 16.0 Advisory Message Test (Manual/Advisory Units)
- 17.0 Pacer Test
- 18.0 SpO2 Monitor Test (for SpO2 Option)
- 19.0 EtCO2 Monitor Test (for EtCO2 Option)
- 20.0 Barometric Pressure Calibration Check
- 21.0 CO2 Accuracy Check (for EtCO2 Option)
- 22.0 NIBP Transducer Calibration Test
- 23.0 NIBP Monitor Test
- 24.0 NIBP Volume Leak Test
- Troubleshooting
- Disassembly Procedures
- Required Equipment
- Safety Precautions
- Removing the Cable Caddy
- Removing the Handle
- Removing the Recorder, AC Charger and Battery Well
- Removing the Front Panel Assembly
- Removing the Side Panels
- Removing the Connector Panel and Bezel
- Removing the ECG Input Connector
- Removing the NIBP Assembly
- Removing the System Brick Assembly
- Disassembling the System Brick Assembly
- Discharging Capacitor
- Front Panel Disassembly
- Replacement Parts
- Functional Description
- Detailed Description
- AC Charger
- SurePower Battery
- Parameter Power Supply (SpO2, EtCO2, NIBP)
- Digital System Board
- Analog System Board
- Pace/Defib Core Engine
- Front Panel Controls
- Peripherals
- Accessories
- Power Management Support Functions
- WiFi
- Interconnect Diagram for the R Series Biphasic Unit
- Delivered Energy at Every Defibrillator Setting into a Range of Loads
- 1.0 Physical Inspection
- 2.0 Front Panel Button Test
- 3.0 Leads Test
- 4.0 Power Supply Test
- 5.0 Leakage Current Test
- 6.0 Paddles Test
- 7.0 Heart Rate Display Test
- 8.0 Pulse Calibration Test
- 9.0 Notch Filter Test
- 10.0 Heart Rate Alarm Test
- 11.0 Defibrillator Self Test
- 12.0 Synchronized Cardioversion Test
- 13.0 Synchronized Cardioversion Test (Remote ECG Monitoring)
- 14.0 Shock Test
- 15.0 Summary Report Test
- 16.0 Advisory Message Test
- 17.0 Pacer Test
- 18.0 SpO2 Monitor Test
- 19.0 EtCO2 Monitor Test
- 20.0 Barometric Pressure Calibration
- 21.0 CO2 Accuracy Test
- 22.0 NIBP Transducer Calibration Test
- 23.0 NIBP Monitor Test
- 24.0 NIBP Volume Leak Test